Loading the Chemistry Protocol Diskette
Description
System Setup
IMMAGE 800 Operations Manual A11403
Page 5-66
September 2016
Loading the Chemistry Protocol Diskette
Description
The chemistry protocol diskette is provided with each IMMAGE 800
Immunochemistry system. It contains essential non-lot specific information about
how to run each chemistry.
When to Load the Protocol Diskette
The chemistry protocol diskette is loaded when the IMMAGE 800 is installed. When
new chemistries become available, a new diskette is provided.
Loading the Protocol Diskette
The instrument status must be in
Standby
in order to proceed with the steps below to
load the chemistry protocols onto the hard drive of the computer.
Step
Action
1
Select
Setup
from the menu bar.
2
Select
<16> Chemistry Protocol Diskette
from the Setup screen.
3
Insert the chemistry protocol diskette into the floppy diskette drive. (Refer
to CHAPTER 2,
System Description
.)
4
Select
<OK>
if the database has been archived.
OR
Select
<Cancel>
to return to the Setup screen without loading the diskette.
5
If the diskette is...
the screen...
successfully read
returns to the
Setup
menu.
unsuccessfully read
displays an appropriate error
message. (Refer to CHAPTER 10,
6
Select
Main
to return to the main display.