Reference Interval Setup
Selecting the Default
System Setup
IMMAGE 800 Operations Manual A11403
Page 5-20
September 2016
Selecting the Default
The default interval and range will be printed when an age is not specified in sample
programming or when age is specified but the reference interval has not been defined
for that age.
Only one default can be chosen for a particular interval and range definition grouped
by chemistry/calculation, sample type and sex.
When the instrument status is in
Standby
, proceed with the steps below to select the
default interval and range.
7
Repeat Steps 1-6 for additional interval and range definitions for the same
chemistry/calculation.
Use the
<Page Up>
or
<Page Down>
buttons to move to other pages of
the same chemistry/calculation.
8
Select another sample type and/or sex to define additional groups of the
chemistry/calculation. Repeat Steps 1-7.
OR
Select
Prev Chem [F9]
or
Next Chem [F10]
to define other chemistries.
Repeat Steps 1-7.
OR
Select any icon from the menu bar to exit the screen.
Step
Action
, continued
(2 of 2)
Step
Action
1
From the Setup screen, select
<4> Reference Interval
2
Select the button beside the desired chemistry/calculation. Multiple
chemistries/calculations may be chosen.
3
Select
<Define Edit>
4
Select a field on the interval/range line that is the chosen default.
5
Select
Select Default [F4]
.
A
">"
symbol will mark the default range on the far left of the line.
6
Select another sample type and/or sex to move to other definitions within
the same chemistry/calculation. Repeat Steps 4-5.
OR
Select
Prev Chem [F9]
or
Next Chem [F10]
to move to other chemistries/
calculations. Repeat Steps 4-5.
OR
Select any icon from the menu bar to exit the screen.