Errors
Error Table
IMMAGE 800 Operations Manual A11403
Utilities
September 2016
Page 10-23
10
E11
Chk Optics
Optics Error, excessive
scatter
1. Dilute and rerun sample.
2. Check and remove bubbles from
reagent cartridge/sample
container.
E12
Chk Optics
Excessive Q Values
Detected
1. Check and remove bubbles from
reagent cartridge/sample
container.
2. Check and remove bubbles from
buffers and diluents.
3. Rerun sample.
E13
Chk Optics
Optics Error Divide-by-
Zero or Log 10 Domain
CALL BECKMAN COULTER
CLINICAL SUPPORT
E14
Chk Optics
Optics Error for AGXS,
Divide-by-Zero or Log
10 Domain
CALL BECKMAN COULTER
CLINICAL SUPPORT
E15
Chk Optics
Optics Error, Bad Data
CALL BECKMAN COULTER
CLINICAL SUPPORT
E16
Chk Optics
Optics Error for AGXS,
excessive scatter
CALL BECKMAN COULTER
CLINICAL SUPPORT
E17
Chk Optics
Optics Error for AGXS,
Bad Data
CALL BECKMAN COULTER
CLINICAL SUPPORT
E18
RGT Temp OR
Reagent Temperature Out
of Range
1. Verify reagent fans are turning.
2. Close reagent cover.
3. Check Status Monitor to verify
reagent temperature is within
range.
E19
Anlg Gnd OR
Analog Ground Out of
Range
CALL BECKMAN COULTER
CLINICAL SUPPORT
E20
Ref Volt OR
Reference Voltage Out of
Range
CALL BECKMAN COULTER
CLINICAL SUPPORT
E21
PSVE PS OR
P5VE Power Supply Out
of Range
CALL BECKMAN COULTER
CLINICAL SUPPORT
E22
P12V PS OR
P12V Power Supply Out
of Range
CALL BECKMAN COULTER
CLINICAL SUPPORT
E23
N12V PS OR
N12V Power Supply Out
of Range
CALL BECKMAN COULTER
CLINICAL SUPPORT
Table 10.6
Displayed or Printed Error Messages
, continued
Error #
Abbreviation/
Message
Explanation
Action
(2 of 21)