As-Indicated Maintenance
Replacing Cuvettes
Utilities
IMMAGE 800 Operations Manual A11403
Page 10-18
September 2016
4
Unscrew cuvette cover plate counterclockwise. Remove cover plate.
Figure 10.14
5
Remove cuvettes by lifting upward. Rotate the wheel as necessary to
access all the cuvettes. Dispose of the used cuvettes in a manner
appropriate for biohazardous materials.
6
Wipe the wheel with lint-free tissue moistened with DI water.
7
Replace all cuvettes, handling by the top edges only. (Refer to Figure
10.15.) Confirm that all cuvettes are seated in the reaction carousel.
Figure 10.15
8
Replace the covers.
9
Select
Utilities
from the menu bar.
10
Select
<7> Wash Cuvettes
from the Utilities dialog box.
OR
Enter
7
in the
Option No.
field and press
[Enter]
.
11
Type the first cuvette number (1) to begin the wash in the
Starting Range
field and press
[Enter]
.
Step
Action
, continued
(2 of 3)