Reference Interval Setup
Before Defining/Editing Intervals/Ranges
IMMAGE 800 Operations Manual A11403
System Setup
September 2016
Page 5-19
5
Before Defining/Editing Intervals/Ranges
If more than one age-defined interval or range is to be entered, it must be entered from
the lowest age range to the highest age range.
Example:
• 1-5 years on the first line
• 6-14 years on the second line
• 15-175 years on the third line
Age ranges cannot overlap.
Example of overlapping age ranges:
• 1-5 years
• 5-175 years
Defining/Editing Intervals and Ranges
The instrument status must be in
Standby
in order to proceed with the steps below to
define reference intervals and critical ranges.
Step
Action
1
From the Reference Interval/Critical Ranges screen,type the low age in the
Low
field. Press
[Tab]
2
Type the age unit in the field next to the low age. Press
[Enter]
.
3
Type the high age in the
High
field. Press
[Tab]
.
4
Type the age unit in the field next to the high age. Press
[Enter]
.
5
Type the low and high reference interval numbers in the
Low
and
High
Reference Interval
fields.
The low reference interval number must be less than the high reference
interval number rounded to the nearest hundredth.
Example:
Acceptable entry: Low-5.06, High-5.07
Unacceptable entry: Low-5.061, High-5.063
6
Optional entry: Type the low and/or high critical range numbers in the
Low
and/or
High Critical Range
fields.
The critical range numbers must be outside of the reference interval.
(1 of 2)