Overview
Introduction
Sample Programming
IMMAGE 800 Operations Manual A11403
Page 7-60
September 2016
Rerunning a Sample
Overview
Introduction
A sample can be rerun when it has a status of
Complete
or
Incomplete
. If the system is
running, a sample programmed to be rerun will be added to the current run. If the
system has returned to a
Standby
status, a new run must be started.
Several options are available when rerunning samples:
• The original sample programs can be edited before rerunning.
• The original sample programs can be rerun.
• Selected chemistries can be rerun for a group of samples.
• The initial dilution made in the dilution well can be reused. The default is to remake
a dilution.
Rerun Results
When tests are rerun:
• The rerun results replace the original results.
• Rerun results are indicated by a flag (R) on the Laboratory Report.
• After a sample is rerun, all results will be collated on the result report.