As-Indicated Maintenance
Replacing Syringes
Utilities
IMMAGE 800 Operations Manual A11403
Page 10-20
September 2016
8
Select
<3> Aspirate volumes
.
9
Enter "125" for the sample syringe or enter "250" for the Reagent Syringe,
select
<OK>
.
10
Unscrew bottom syringe screw (1) and top syringe screw (2). (Refer to
Figure 10.16.)
Figure 10.16
11
Remove syringe for replacement.
Figure 10.17
12
Insert new syringe.
13
Select
Main
from the menu bar.
14
After the status returns to
Standby
, prime the system and observe syringes
for leaks or bubbles. Refer to Daily Maintenance Procedures, "Priming,"
in this chapter.
15
Run controls to verify system performance.
Step
Action
, continued
(2 of 2)