Calibrating a UDR Chemistry
Running the UDR Calibration
IMMAGE 800 Operations Manual A11403
System Setup
September 2016
Page 5-99
5
Figure 5.44
User-Defined Calibration Load List Screen
Running the UDR Calibration
The instrument status must be in
Standby
in order to proceed with the steps below to
run a UDR calibration.
E011422S.EPS
Step
Action
1
Load the calibration samples in the appropriate racks.
2
Select
Main
from the menu bar and
Run
.
3
Select
<OK>
in the Check Dilution Segments dialog box to start the run.
OR
Select
<Cancel>
to exit without starting the run.
4
When the calibration run is finished, a multi-point or single-point UDR
Calibration Results report is printed. Continue to "Approving a UDR
Calibration."