Checking Dilution Segment Status and Clearing Dilution Segments
After Clearing Segments
IMMAGE 800 Operations Manual A11403
System Status/Instrument Commands
September 2016
Page 11-3
11
Figure 11.1
Dilution Segments Dialog Box
After Clearing Segments
Always replace the cleared dilution segments with unused segments on the sample
carousel before a run is started.
Replacing Dilution Segments
The instrument status must be in
Standby
in order to proceed with the steps below to
replace dilution segments.
Additional Information
for more information on Dilution
Segments status.
E010293S.EPS
Step
Action
1
Lift the cover of the sample carousel.
2
Rotate the sample carousel by pressing the advance button in order to
access the desired dilution segment, if necessary.
3
Lift the segment to be replaced off the sample carousel.
4
Dispose of the used segment in a manner appropriate for biohazardous
materials.
5
Place an unused segment in the empty position on the sample carousel.
6
Repeat steps 2-5 until all desired segments are replaced.