Loading UDR Reagent Cartridges
Loading UDR Cartridges
System Setup
IMMAGE 800 Operations Manual A11403
Page 5-88
September 2016
Figure 5.37
UDR Cartridge Refill Screen
The UDR Cartridge Refill screen (Figure 5.37) is defined in Table 5.13:
7
Select
<Refill>
. The system will update tests remaining in the cartridge to
the number of tests defined in the Defined/Edit-User Defined Chemistry
screen.
OR
Select
<Exit>
to return to the Defined/Edit-User Defined Chemistry
screen without refilling the cartridge.
8
Refill the cartridge.
Table 5.13
UDR Cartridge Refill Screen
Screen Item
Screen Definition
Chem Name
Chemistry Name as defined in Figure 5.32.
Reagent Lot Number
Reagent Lot Number as defined in Figure 5.32.
Tests per cartridge
Tests per cartridge as defined in Figure 5.32.
Serial Number
Cartridge Serial number as defined in Figure 5.32
Tests left
Number of tests remaining from tests per cartridge
defined in Figure 5.32.
Total tests left
Number of tests remaining from 300 total tests per
cartridge.
Step
Action
, continued
(2 of 2)
E011420S.EPS