Loading/Clearing Buffers and Diluents
Clearing a Buffer or Diluent Position
Reagents/Calibration
IMMAGE 800 Operations Manual A11403
Page 6-12
September 2016
Clearing a Buffer or Diluent Position
The instrument status must be in
Standby
in order to proceed with the steps below to
clear a buffer or diluent position.
Step
Action
1
Select
Rgts/Cal
from the menu bar.
2
Select
Buffer/Diluent [F3]
.
3
Open the reagent or sample compartment cover.
4
Remove the buffer or diluent from the carousel.
5
Close the reagent or sample compartment cover.
6
Select the options button
<
▼
>
beside the position to be cleared.
7
Select
<Clear>
to clear the position.
OR
Select
<Cancel>
to return to the Reagent Status/Calibration Status screen
without clearing the position.