Programming a Control
Programming a Non-bar Coded Control
Sample Programming
IMMAGE 800 Operations Manual A11403
Page 7-40
September 2016
Programming a Non-bar Coded Control
Follow the steps below to program a non-bar coded control.
Step
Action
1
Select
Select Control [F5]
from the Program Sample screen.
2
Type the number of the desired control and press
[Enter]
OR
Select the number beside the desired control.
OR
Select
<Cancel>
to return to the Program Sample screen without selecting
a control.
3
Enter a Rack and Position number. (Refer to Figure 7.11.)
AND
Select the options button
<
▼
>
beside the Control ID field to select a
Control ID. (A Control ID is optional.)
NOTICE
If a control is programmed with a Control ID and the host sends a
patient sample with an ID identical to the control, the programmed
control will be deleted.
4
Select or deselect panels and/or chemistries if necessary.
Select
Clear Chems [F7]
to deselect all chemistries.
(Refer to Selecting Chemistry Tests by Panel and Selecting Chemistry
Tests by Chemistry in this Section.)
5
Select
Sample Options [F3]
if needed.
(Refer to Selecting Sample Options in this Section.)
6
Select
Select Control [F5]
to return to the Select Control dialog box
without saving the control program.
OR
Select
Save [F10]
to save the control program and return to the Program
Sample screen.
OR
Select
Cancel Edit [F9]
to return to the Program Sample screen without
saving the control program.