Re-Enabling Calibration
Introduction
IMMAGE 800 Operations Manual A11403
Reagents/Calibration
September 2016
Page 6-35
6
Re-Enabling Calibration
Introduction
The previous successful calibration can be re-enabled only when a calibration fails.
Re-enabling Calibration
The instrument status must be in
Standby
in order to proceed with the steps below to
re-enable a calibration.
After Calibration is Re-enabled
• The previous calibration is re-enabled for all reagent cartridges of the same lot.
• Results generated by a re-enabled calibration are flagged on the laboratory and
patient chartable reports under the Instrument Code. (Refer to APPENDIX A,
Step
Action
1
Select
Rgts/Cal
from the menu bar.
2
Select the options button
<
▼
>
to the right of the Cal ID of the failed
calibration.
3
From the Re-enable Previous Calibration dialog box, select
<OK>
to re-
enable the calibration.
OR
<Cancel>
to return to the Reagent Status/Calibration Status screen
without re-enabling the calibration.