Calibrate Touch Screen
Introduction
IMMAGE 800 Operations Manual A11403
Utilities
September 2016
Page 10-67
10
Calibrate Touch Screen
Introduction
The IMMAGE 800 system can access screen components with the optional touch
screen monitor. After calibration of the touch screen, software functions can be
activated by the touch of a finger or other object. Any icon, button, or field on the
screen can be selected by touching the icon, button, or field on the screen.
Calibrating the Touch Screen
The instrument status must be in
Standby
in order to proceed with the steps below to
calibrate the touch screen.
Step
Action
1
Select
Utilities
from the menu bar.
2
Select
<11> Calibrate Touch Screen
from the Utilities dialog box.
OR
Enter
11
in the
Option No.
field and press
[Enter]
3
To calibrate and enable the touch screen, with one finger or an appropriate
object, touch the crosshairs in the upper, left quadrant of the display area
twice. Next, touch the crosshairs in the lower, right quadrant of the display
area twice.
OR
To exit without saving the calibration, select
<Cancel>
or any key on the
keyboard.
NOTICE
The touch screen will not be activated if the first calibration attempt is
canceled prior to touching both crosshairs. If a previous calibration
exists, canceling a calibration attempt retains the previous calibration
data.
4
After calibration, software functions can be activated by the touch of a
finger or other object. Any icon, button, or field on the screen can be
selected by touching the icon, button, or field on the screen. Once the
touch screen is activated, it cannot be turned off.