Calibration History
Introduction
IMMAGE 800 Operations Manual A11403
Reagents/Calibration
September 2016
Page 6-21
6
Calibration History
Introduction
The Calibration History window allows the operator to review the previous four
calibrations and displays the following information:
• chemistry name
• reagent lot number
• date
• time
• Cal ID
• scale factor
• reaction buffer
• sample diluent
Displaying Calibration History
Calibration history is not available for User-Defined Reagents. Up to four successful
calibrations will be displayed. Failed calibrations will not be displayed.
Re-Enable Calibration
If the most recent calibration failed and there was a previous successful calibration, the
user has the option to re-enable the previous calibration.
Step
Action
1
Select
Rgts/Cal
icon.
2
From the Reagent Status/Calibration Status screen, select the pull-down on
the
Cal ID
field to display the Calibration History window. Refer to Figure
3
The calibrations display ascending date/time order, with the oldest
calibration listed first.
4
Select
<OK>
to return to the Reagent/Status/Calibration screen.
Step
Action
1
From the Reagent Status/Calibration Status screen, select Re-enable
Calibration. The calibration status of the test changes to Cal Re-enable.
2
The calibrations display ascending date/time order, with the oldest
calibration listed first.
3
Select
<OK>
to return to the Reagent/Status/Calibration screen.