Daily Maintenance Procedures
Cleaning Instrument Parts
Utilities
IMMAGE 800 Operations Manual A11403
Page 10-10
September 2016
Cleaning Instrument Parts
The instrument status must be in
Standby
in order to proceed with the steps below to
clean the reagent/sample probes and mixer paddles.
while priming, or
during routine
operations, fluid leaks
or bubbles are present,
remove bubbles from reagent (1) and sample (2) syringes,
syringe lines (3), tips (4), and shear valves (5). (Refer to
Figure 10.3) Bubbles may be removed by gently tapping on
the tubing or syringe during the prime function.
Refer to As-Indicated Maintenance in this section for
syringe replacement.
Figure 10.3
If...
then...
, continued
(2 of 2)
Step
Action
1
Raise reagent/sample crane probe covers as shown.
Figure 10.4
(1 of 2)