background image

Instruction Manual

HI 8314

Portable pH/mV/ºC 

meter

w w w . h a n n a i n s t . c o m

Hanna Instruments reserves the right to modify the design, construction

and appearance of its products without advance notice.

WARRANTY

All Hanna Instruments

 meters are warranted for two years 

against

defects in workmanship and materials when used for their intended pur-
pose and maintained according to instructions. 

The electrodes and the

probes are warranted for a period of six months. 

This warranty is

limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed main-
tenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of purchase,
serial number and the nature of the failure. If the repair is not covered by
the warranty, you will be notified of the charges incurred. If the instrument
is to be returned to Hanna Instruments, first obtain a Returned Goods
Authorization Number from the Customer Service department and then
send it with shipment costs prepaid. When shipping any instrument, make
sure it is properly packaged for complete protection.

All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner.

Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at

 [email protected]

.

This instrument is in compliance with the 

 directives.

Remove the instrument from the packing material and exam-
ine it carefully to make sure that no damage has occurred
during shipment. If there is any damage, notify your Dealer.
The meter is supplied complete with:
• 

HI 1217D

 pH electrode

• calibration and cleaning solution sachets
• 9 V battery, calibration screwdriver and instructions.

Note:

Conserve all packing material until the instrument has
been observed to function correctly. Any defective item
must be returned in its original packing.

PRELIMINARY EXAMINATION

GENERAL DESCRIPTION

HI 8314

 is an hand-held pH/mV/°C meter designed for to be

rugged, easy-to-use, reliable and practical.
It is ideal for education and for measurements on the field.
The pH, mV and °C ranges are easily selected using the
membrane keyboard on the front panel.
Calibration procedure is very simple and adjustments are
easily made with two trimmers on the front panel.
The built-in temperature sensor in the pH electrode allows
automatic temperature compensation of pH readings.
Moreover, 

HI 8314

 features a low battery indicator that

alerts the user when the bettery needs to be replaced.

FUNCTIONAL DESCRIPTION

SPECIFICATIONS

1) DIN connector for pH or ORP electrode
2) Liquid Crystal Display
3)

mV

 key, to display the mV (ORP) readings when using

an ORP electrode or the mV equivalent to pH values
when using a pH electrode

4)

STD 

trimmer

 

for OFFSET calibration of pH

5)

SLOPE 

trimmer for SLOPE calibration of pH

6)

°C

 key, to display the temperature measurement

7)

pH

 key, to display the pH value

8)

ON/OFF 

key, to turn the meter ON and OFF

Range

0.00 to 14.00 pH

±1999 mV

0.0 to 100.0 °C

Resolution

0.01 pH / 1 mV / 0.1°C

Accuracy 

(@20°C/68°F)

±0.01 pH

±1 mV

±0.4°C

Typical EMC Deviation

±0.07 pH

±5 mV

±1°C

pH Calibration

Manual, 2 point, through trimmers

Offset Calibration

±1 pH

Slope Calibration

85 to 105%

Temperature Compensation

Automatic, 0 to 70°C (32 to 158°F)

Electrode (*)

HI 1217D

 (

included

)

Battery Type

1 x 9V alkaline

Battery Life

100 hours of continuous use

Environment

0 to 50°C (32 to 122°F);

RH max 95% non-condensing

Dimensions

164 x 76 x 45 mm (6.5 x 3.0 x 1.8")

Weight

180 g (6.3 oz.)

(*)

 The max. operating temperature of the 

HI 1217D

 pH electrode is 80ºC. To

perform measurements at higher temperature use adequate electrodes.

DISPLAY CODE GUIDE

°C

symbol indicates the meter is in temperature mode.

pH

symbol indicates the meter is in pH mode.

mV

symbol indicates the meter is in mV mode.

Recommendations for Users

Before using these products, make sure that they are entirely suitable for the environment in
which they are used. Operation of these instruments in residential area could cause unaccept-
able interferences to radio and TV equipments, requiring the operator to take all necessary
steps to correct interferences. Any variation introduced by the user to the supplied equipment
may degrade the instruments' EMC performance. To avoid damages or burns, do not perform
any measurement in microwave ovens.