Clearing a Sample by Rack and Position
Clearing by Rack and Position
Sample Programming
IMMAGE 800 Operations Manual A11403
Page 7-78
September 2016
Clearing a Sample by Rack and Position
Clearing by Rack and Position
Follow the steps below to enter Rack(s) and Position(s) to be cleared.
Step
Action
1
Select
Clear Samples [F7]
from the Program Sample screen.
2
Type the rack number(s) desired to be cleared in the
Rack(s)
field (up to 15
characters). Numbers can be separated by a comma as a series and/or by a
dash as a range.
3
If...
then...
one rack number is
entered,
Press
[Enter]
to access the
Pos(s)
field.
Type the position numbers desired to be
cleared (up to 15 characters). Numbers can
be separated by a comma as a series and/or
by a dash as a range. (
Example
: 1,2,5-8)
Note: If one rack number and no position
number is entered, all positions on the rack
are automatically selected.
more than one rack
number is entered,
• All positions on each rack are
automatically selected.
• The
Pos(s)
field cannot be accessed.
4
Select
<OK>
.
Select
<OK>
again to confirm the Racks and Positions or
<Cancel>
to
return to the Clear Samples dialog box.
OR
Select
<Cancel>
to return to the Program Sample screen without clearing