Calibrating a UDR Chemistry
Clearing a UDR Cal Rack
IMMAGE 800 Operations Manual A11403
System Setup
September 2016
Page 5-101
5
Clearing a UDR Cal Rack
The instrument status must be in
Standby
in order to proceed with the steps below to
clear a UDR Cal rack.
Clearing the rack used for a UDR Single Point Calibration Update before approving it
will also clear the previous UDR calibration (Multipoint or Single Point Update).
Therefore, if you choose to repeat a Single Point Calibration Update, do it without
clearing the current calibration rack.
Figure 5.46
Clear UDR Calibration Racks Dialog Box
NOTICE
Do not clear a UDR Cal rack until the final curve-fit model for the calibration has been
selected.
Step
Action
1
From either the User-Defined Chemistries screen or the UDR Cal Assign
screen, select
Clear UDR Cal Rack [F8]
2
Enter the racks to clear in the
Rack(s)
field.
3
Select
<OK>
to clear the rack.
OR
Select
<Cancel>
to exit without clearing.
NOTICE
Clearing the User-Defined Calibration Rack will clear the rack positions.
The calibration results for this rack will be deleted. After clearing the
rack, the calibration results cannot be displayed with the current model,
with a different model, or be printed.
E014055S.EPS