Sample and Reagent Crane Probe
Replacing Sample and Reagent Crane Probe
IMMAGE 800 Operations Manual A11403
Utilities
September 2016
Page 10-73
10
3
With the probe over the wash station (1), remove the tubing from the top of
the probe assembly by unscrewing the black fitting (2). (Refer to Figure
10.37.)
Figure 10.37
4
Locate the access cover (1) for the probe level sense cable (2) by tracing
the cable from the probe.
Remove the access cover using a Phillips screwdriver. (Refer to Figure
10.38.)
Figure 10.38
Step
Action
, continued
(2 of 6)
1
A012185P.EPS
2
1
A014084P.EPS
2