Requesting and Canceling Calibration
Checking and Clearing Racks
Reagents/Calibration
IMMAGE 800 Operations Manual A11403
Page 6-28
September 2016
3
If...
then...
one rack number is entered, Press
[Enter]
to access the
Pos(s)
field.
Type the position numbers desired to
clear (up to 15 characters). Numbers
can be separated by a comma as a series
and/or by a dash as a range.
Note: If one rack number and no
position number is entered, all
positions on the rack are automatically
cleared.
more than one rack number
is entered,
• all positions on each rack are
automatically cleared.
• the
Pos(s)
field cannot be accessed.
4
Select
<OK>
to clear the racks and positions.
OR
Select
<Cancel>
to return to the Reagent Status/Calibration Status screen
without clearing.
5
If the calibrator sample
is...
then...
bar coded
Select
Save [F9]
to save the request.
OR
Select
Cancel [F10]
to return to the
Reagent Status/Calibration Status
screen without saving.
not
bar coded
Go to Assigning Calibrator Rack and
Position in this section.
Step
Action
, continued
(2 of 2)