Errors
Error Table
IMMAGE 800 Operations Manual A11403
Utilities
September 2016
Page 10-27
10
E63
ORLO
Unstable Reaction
(excluding Drugs)
Refer to IMMAGE Immunochemistry
Systems,
Chemistry Information
Manual
for appropriate measuring
range.
E64
RXN Error
Unstable Reaction
(Drugs)
1. Verify all constituents (reagents,
samples, buffer, diluents, and
segments) are available.
2. Rerun sample.
E65
ORHI
High Result
Refer to IMMAGE Immunochemistry
Systems,
Chemistry Information
Manual
for appropriate measuring
range.
E66
RXN Error
Unstable Reaction
CALL BECKMAN COULTER
CLINICAL SUPPORT
E70
Invalid Cal
Invalid Calibration Type,
Incorrect Reagent Card or
Bad Read
1. Reread reagent card.
2. Reread calibration card.
E71
Cal Reps
Too Few Calibrators:
Need Two or More
Replicates for
Calculation (Only one
Replicate Calculation)
Repeat calibration.
E72
No Param
No Valid Reagent
Parameters A-G, Reagent
Card with no Parameters
or Bad Read
Reread reagent card.
E73
RGT Handler
Reagent Handler Error
CALL BECKMAN COULTER
CLINICAL SUPPORT
E76
Loop Error
Out of range result failed
to reach final dilution
level
Rerun pending test.
E77
DilWell LVL
Level sense failure on a
dilution well
Rerun sample.
E79
S/W Error
No memory available
Reboot Instrument.
E80
Bad Data
Bad Data
CALL BECKMAN COULTER
CLINICAL SUPPORT
Table 10.6
Displayed or Printed Error Messages
, continued
Error #
Abbreviation/
Message
Explanation
Action
(6 of 21)