Calibration Guide 

 

T51 Indicators

 

Ohaus Corporation  www.ohaus.com 

 

i

 

T51 Indicators

 Calibration Guide 

T51 Indicators have three calibration processes: 

Zero

Span

, and 

Linearity

 Calibration.  

Follow the prompts for calibration mass values. 

  Ensure that the scale is level and stable during the entire calibration process.  

  Allow Indicator to warm up for five minutes after stabilizing to room temperature.  

  If the Calibration Menu is locked, the scale may be set in Legal For Trade mode.  

  To abort calibration, press the 

Exit 

button anytime during the calibration process. 

Zero Calibration 

 

Zero calibration uses one calibration point. This point is established with no weight on the scale. 
Use this calibration method to adjust for a different pre-load without affecting the Span or 
Linearity calibration.  

Press and hold 

Menu

 until 

C.A.L. 

appears, then press 

Yes

.   

(

LFt.NO

 appears briefly before 

C.A.L.

, if the Indicator is not in LFT mode.) 

ZErO

 appears.  Press 

Yes 

to initiate Zero Calibration. 

 

The display flashes 

and the calibration unit. Press 

Yes 

to establish the 

zero point. 

 

The display shows 

--C--

 while the zero point is established. 

 

When zero calibration is completed, the display shows 

dONE

 

Then the scale exits to the active weighing mode and displays the actual 
weight value. 

ã

 

Span Calibration 

Span Calibration uses two points to adjust the Indicator.  The first point is the zero value, where 
there is no weight on the scale.  The second point is the Span value, where a calibration mass 
is placed on the scale. 

Press and hold 

Menu

 until 

MENU 

appears.  Press 

Yes

(

LFt.NO

 appears, then 

C.A.L.

, if the Indicator is not in LFT mode.) 

Press 

Yes

.  

SPAN

 appears.  Press 

Yes 

to initiate Span Calibration. 

 

The display flashes the span calibration point.  Place the specified weight 
on the scale and press 

Yes

.  

(Press 

No 

to increment the selections, or 

Back 

to decrement the selections.  

When desired value appears, place specified weight on scale; press 

Yes

.) 

 

The display shows 

--C--

 while the span point is established. 

 

When the span point is established, the display flashes 

0

.  

Remove the weight from the scale; press 

Yes 

to establish the zero point. 

 