TM50 Setup Procedure 

This document outlines how to setup your TM50 for automated monitoring using GeoMoS. 

1.

 

Ensure that GeoMoS is not trying to control the instrument you are setting up during this process. 

The best way to ensure this is to stop GeoMoS from running or de-activate GeoMoS in the 

“Sensor Manager” before you begin. 

2.

 

Screw the tribrach down on the 5/8

th

 thread that protrudes from the top of the pillar through the 

hole in the aluminum instrument housing. Ensure it is screwed down tightly as this is what locks 

the instrument housing to the pillar. 

3.

 

Once it is fitted tightly, use the foot screws on the tribrach to make it level with the bubble built 

in to the tribrach. 

4.

 

Now place the instrument into the tribrach and use the locking clamp to lock it down. Check that 

you can access the Lemo connecter in the TM50 when you do this and then connect the Leica 

cable aligning the red dots. 

5.

 

 Turn the TM50 on by pressing and briefly holding down the power button on the keyboard which 

looks like this.

 

6.

 

Now you must fine tune the TM50 using the electronic bubble. This is more precise than the 

bubble on the tribrach. The arrows on the screen show the direction to turn the foot screws.  

Once it is perfectly level, select “Next”. See below. 

 

7.

 

Next you will be prompted to enter the atmospheric conditions. This is not necessary as GeoMoS 

uses a dedicated weather station however, you will need to ensure that the atmospheric ppm” is 

set to zero.

  

 