background image

 

Oil Intensifier System 

P/N 401746

 

Instruction Sheet

 

 

1.  SUPPLY connection to OPG1 or MPG1 TEST Port (DH500) 

5.   INTENSIFIER SHUTOFF valve 

2.  TEST connection to PG7302 (DH500) 

6.  TEST connection to Device Under Test (DUT) 

3. Intensifier 
4.   LOW PRESSURE SHUTOFF valve 

7.  Safety Head with 120 MPa (17 500 psi)  
 

Rupture Disc and Muffler 

 

DESCRIPTION 

The Oil Intensifier System (OIS) is intended to generate 
pressure up to 500 MPa (75 000 psi) using an OPG1 or 
MPG1 (whose maximum pressure is 200 MPa (30 000 psi)). 
The OIS includes a 5:1 intensifier. The intensifier is used 
only when necessary to generate pressure > 200 MPa 
(30 000 psi). 

INSTALLATION 

n

  Position the OIS between the OPG1 or MPG1 and 

the PG7302 with the PG7302 to the left. 

o

  Using a DH500 coned and threaded tube, connect 

the OPG1 or MPG1 TEST port to the OIS SUPPLY 
port (1). 

p

  Using a DH500 coned and threaded tube or two 

tubes and a tee, connect the PG7302 TEST port to 
the OIS lateral TEST port (2). 

Operation:  Intensifier Not in Use (Pressure  

 200 MPa (30 000 psi))

 

n

  Open LOW PRESSURE SHUTOFF valve (4) and close 

INTENSIFIER SHUTOFF valve (5). In this condition, 
OPG1 or MPG1 generates pressure straight through 
the Intensifier System, by passing the intensifier. 

o

  Plug TEST port (6) if it is not being used. 

Operation:  Using Intensifier (Pressure > 200 MPa 
(30 000 psi))

 

OPG1 or MPG1 generates and adjusts pressure up to 
200 MPa (30 000 psi).  

Generally, the OIS intensifier is 

used only when running tests that include pressures 
greater than 200 MPa (30 000 psi

). The OIS may be 

switched into and out of use under pressure. 

n

  When running tests that include pressures above  

200 MPa (30 000 psi), connect the device under test 
(DUT) to the upper TEST connection (6) of the OIS. 

o

  Set pressures under 100 MPa (15 000 psi) using the 

OPG1 or MPG1, without the intensifier. 

p

  To switch in the intensifier, first close the LOW 

PRESSURE SHUTOFF valve (4), then open the 
INTENSIFIER SHUTOFF valve (5).  In this condition the 
OPG1 or MPG1 pressure is shut off from the PG7302 
and is applied to the low side of the intensifier. 

 

Increasing pressure on the low side of the intensifier 
results in pressure changes five times greater on the 
high side (PG7302 side) of the intensifier. 

q

  Use the OPG1 or MPG1 to generate and adjust 

pressure as necessary to set test points and float the 
PG7302 piston. 

 

 

 

Reviews: