Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-56
1.
Remove the rear cover of the C-Arm. Release the three screws and open the package box. The
parts are on the back of the package.
2.
Disconnect all cables connected to the power supply. Loose the screws and replace a new part.
3.
Connect the cables according to the labels. Reassemble all parts.
4.
Power on the system. Verify that the Workstation and C-Arm complete software boot and
initialization with no errors reported on their displays.
DC bus capacitor (P/N: 5329278)
1.
Remove the rear cover of the C-Arm. Release the three screws and open the package box. The
part is on the back of the package.
There are two capacitors here but only one in FRU kit. Please identify the trouble for the part to determine
the quantity of FRU parts.
MISC, DC POWER
5V,
+15V,-15V AND
+24V (P/N:
5085800)
MISC, DC POWER
5V, +12V
AND +24V (P/N:
5085801)
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...