Installation
3-45
2.
Place X-ray Keyswitch in XRAYOFF position (
). Press Handswitch and verify that no X-rays
are generated. Press vertical lift button and verify the vertical lift column is operational.
3.
Place X-ray Keyswitch in ON position (
). Press Handswitch, footswitch and vertical lift button
to verify X-ray and vertical lift column are enable.
Emergency Switch
Perform the following test to verify the emergency switch.
1.
Press the emergency switch during raising C-Arm.
2.
Verify that vertical column operation and exposure are disabled.
3.
Release the emergency switch. Verify that the vertical column operation and exposure are
available.
3.6.3.
Auto Technique Tracking
Select Auto Fluoroscopy mode (
)
, and the NORM field size to take an exposure. Put the copper
filter on the Monoblock head. Verify that the kVp changes accurately as copper filters are inserted into
the field in the following ranges:
Model
Copper Filter Thickness
1mm
2mm
3mm
Brivo OEC 715/785
(I.I PN: 5075001):
kVp range
64
±
4kVp 74
±
4kVp 82
±
4kVp
Brivo OEC 715/785/865
(I.I PN: 5085917):
kVp range
62
±
4kVp 72
±
4kVp 80
±
4kVp
Brivo OEC 865 (I.I PN: 5075008):
kVp range
62
±
4kVp 72
±
4kVp 80
±
4kVp
If the kVp is not in the range, refer to 4.6.7 for calibration.
Emergency Switch
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...