Installation
3-3
3.2.3.
Power Supply Requirement
Voltage: AC 100V/110V/120V|200V|220V/230V/240V
±
10%
Frequency: 50Hz/60Hz
±
3Hz
Rated Momentary Line Current (shorter than 5s):
•
30A (100V/110V/120V)
•
28A (200V)
•
25A (220V/230V/240V)
Rated Continuous Line Current:
•
20A (100V/110V/120V)
•
12A (200V)
•
10A (220V/230V/240V)
Maximum line impedance: 0.3
Ω
|0.6
Ω
|0.6
Ω
Max Line Regulation (Percent): 15% at maximum radiographic exposure.
Line regulation is based on measurements made at the input (primary winding) of the isolation
transformer under standby conditions and at maximum radiographic exposure by the following equation:
Line regulation = 100 (Vn - V1)/V1
Where Vn = no load voltage (standby), and V1 = loaded voltage (maximum radiographic exposure).
16A power supply is required in China. And 20 A is required in Japan.
Suitable sockets are required both in Scan room and Operating room for maintenance.
Power supply should comply with requirement of the system power supply requirements.
The sockets should be 3 cores with protective earth, which complies with the requirement
of the applicable electrical codes. Other equipment electrically connected with this
apparatus shares the same grounding connection.
You should check the local power supply first. If it is not 200 V/220 V/230 V/240 V (50 Hz),
you need change the power supply cables and change the connections of the system.
Refer to the Power Supply Adaptation section3.5.1 for details.
3.2.4.
Environments
Operating Conditions:
Temperature: +10
°
C~+40
°
C
(
+50F~+104F
)
Humidity: 20%~80% Relative, non-condensing
Atmospheric Pressure: 700hPa~1060hPa
Transportation and Storage:
Temperature: -20
°
C~+55
°
C (-4F~+131F)
Humidity: 10%~95% Relative, non-condensing
Atmospheric Pressure: 500hPa~1060hPa
WARNING
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...