Replacement
7-27
3.
Remove the plate of the wheel. Loose the
screws and then remove the cable pusher.
4.
Replace the new parts in steering none brake position and reassemble all the other parts.
5.
Move the C-Arm and verify that the C-Arm can move smoothly.
Wheels of Workstation
Brivo OEC 715/785 and Brivo OEC 865 use different Workstation wheels. Brivo OEC 715/785 use universal
wheels with brake. Brivo OEC 865 use central control wheel and non-central control wheels.
For Brivo OEC 715/785 (PN: 5343601)
1.
Lift up the side of the Workstation carefully and keep the balance of the Workstation.
2.
Remove the down side cover secure screws and remove the down side covers.
3.
Remove the workstation wheel by loose the nut.
4.
Assemble wheel as above reverse steps.
5.
Move the Workstation and verify that the Workstation can move smoothly.
For Brivo OEC 865
Central Control Wheels (5075965)
1.
Lift up the side of the Workstation carefully and keep the balance of the Workstation.
2.
Loose the screws and remove the down side cover.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...