Calibration
4-23
Good Image:
ABC mode, manual exposure
mode, in 60KV, 1.1mA, there is
NO
black margin
in circular mask ring
Good Image:
None_ABS and None_ABC,
manual exposure in 60KV, 1.1mA with contrast
in 65 and rightness in 40, there is
NO
black
margin in circular mask ring.
Difference of two reading scales between display mark and light image margin in image
area 1/3, for Brivo OEC 865 system shall be
≤
1scale of display mark
, for Brivo OEC
715/785 shall be
<6 scales of display mark.
Difference of two reading scales between display mark and light image margin in image
area 2/4, for Brivo OEC 865 system shall be
≤
1scale of display mark
, for Brivo OEC
715/785 shall be
<6 scales of display mark.
Good
Good
Circular Mask
Circular Mask
1
2
3
4
1
2
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...