Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-28
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0304520
Warning
Init MWL server
list failed
Connectivity configuration
loading failure. Worklist server
list missed. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0304521
Warning
Load anatomic
region
configuration
file failed
Anatomic Regions list missed.
Please restart the system. If
this issue persists, please call
for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0304523
Warning
Init date & time
format failed
Setup configuration loading
failure. Date/Time format
setting missed. Please restart
the system. If this issue
persists, please call for service.
0304524
Warning
Get string failed Loading message fail. Please
restart the system. If this issue
persists, please call for service.
0304525
Warning
Stop
ExamManagem
ent failed
Stop ExamManagement failed.
Please restart the system. If
this issue persists, please call
for service.
0304526
Error
ExamManagem
ent module
register OSGI
service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0304527
Error
ExamManagem
ent module get
OSGI service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0304528
Error
ExamManagem
ent module
unregister OSGI
service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0304529
Warning
Stop
application
exception
Close ExamManagement
exception. Please restart the
system. If this issue persists,
please call for service.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...