Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-68
After removing
Monoblock, collimator and Image
Intensifier, C-cam will in unbalance status
. The
back of C-arm will lower than normal. Take care
anything would be pressured by C-arm. Find wood
block to support the mainframe is appreciated.
9.
Let C-arm in Up/Down position and Down/Up position to make
the cable go through the C-arm.
10.
Install Monoblock support, monoblock
11.
collimator and Image intensifier.
12.
Connect relative cables, install the new bundle cable sleeves and cable supports, route all the cables
and install all the other parts which were removed on above steps.
13.
Power on system to ensure system boot up successfully.
14.
Refer to 7.3.2 calibration matrix and chapter 4 to do calibration and checks.
15.
After calibration and check, shut down system and reinstall covers removed in above steps.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...