1.
Power off system and refer to 7.3.4 remove Workstation back, side, front covers
2.
Disconnect the cables related to system interface board
3.
Release 4 secure screws under cabinet, and then pull out cabinet from Workstation frame.
4.
Release 5 screws which secure System Interface Board and replace a new one.
5.
Connect cables and assemble the cabinet.
6.
Assemble Workstation front, side covers.
7.
Power on system and check the system can boot up functionally.
Mainframe Control Panel Kit (including membrane, display board and support board) (P/N:
5075907)
1.
Remove the rear cover of the C-Arm.
2.
Disconnect all cables connected to control panel.
3.
Loose the screws that secure the control panel and remove the metal shield strips. Then remove
the control panel.
4.
Replace a new control panel and connect the cables. Install the C-Arm cover.
5.
Power on the system. Pressing the buttons and verify that relative indicator can light.
Workstation Control Panel Kit (5075908)
1.
Remove all screws that secure the front upper cover.
Cabinet
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...