Installation
3-17
3.
Workstation Inspection
Inspect each workstation component listed under the heading workstation inspection. Select one check
box for each item in the list.
WorkStation Inspection list:
Physical Location
No problem
found
Loose or missing
parts
Broken or
damaged parts
Chipped or
scratched paint
Monitor and Arm
Keyboard
Mouse
Covers
Frame
Handles
Wheels
3.4.3.
Internal Inspection
Observe electrostatic discharge safety procedures.
Removing any covers exposes dangerous electrical voltages on several components.
When possible, disconnect system from AC line before removing any covers. When it is
necessary to leave the power applied, use extreme care to avoid death or injury from
electrical shock.
1.
Remove workstation cover. Inspect the internal labels (refer to chapter 1.2.11) and parts. Check if
there are loose parts.
2.
Remove the C-Arm’s covers. Check all cables and PCBs inside C-Arm. Look for loose PCBs and
connectors and reset.
WARNING
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...