Calibration
4-11
5.
mA accuracy in Radiography Mode:
Measure the voltage of TP7 (MA-SMP) on MCB board with oscilloscope:
mA measured=Sample Voltage (TP7) (unitV)*1000/200ohm
mA real=mA accuracy measure
mA accuracy=[(mA set – mA actual )/mA set]*100%
Set the exposure mode and parameter to check to the voltage (TP7) values. The voltage of TP7 should be
in the range of the form below. Otherwise, mA calibration procedures should be performed until all the
voltage values meet the requirement in the list.
For power input: 100V/110V/120V (
Setting in Exposure Management
)
Exposure mode KV and mAs value setting
mA value
Voltage (TP7)
minimum (V)
Voltage (TP7)
maximum (V)
Radiography
40KVp@40mAs
10mA
-2.20
-1.80
110KVp@40mAs
10mA
-2.20
-1.80
For power input: 200V/220V/230V/240V (
Setting in Exposure Management
)
Exposure mode KV and mAs value setting
mA value
Voltage (TP7)
minimum (V)
Voltage (TP7)
maximum (V)
Radiography
40KVp@80mAs
20mA
-4.40
-3.60
110KVp@40mAs
20mA
-4.40
-3.60
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...