Calibration
4-17
12.
Click “Ok” to finish the calibration.
13.
Power off the system, then remove the extend board.
14.
Power on the system and perform Preheat mA Calibration.
mA Preheat Table Calibration:
Provide appropriate X-ray protection on surface of Image Intensifier in calibration and
verification to protect Image Intensifier.
15.
Short J5 on MCB board. Select “Preheat mA calibration” in calibration screen of service mode.
“CA21” will display on control panel. Press footswitch or hand switch and hold on until “END END”
is displayed on control panel LED field, which would last for about 20 minutes.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...