Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
3-46
3.6.4.
Image Resolution
Place spatial resolution tool in the front of the image intensifier. The angle between the horizontal axis of
the image and the resolution tool must be 45
°
. Select appropriate kV value. Take an exposure and stop
once image is stable. The image is frozen on the left monitor. Verify the resolution in Lp/mm is equal to,
or greater than, the values listed below. Repeat and select any of three field sizes (NORM, MAG1 and
MAG2) to verify the resolution.
Field Size
Model
MAG NORM
MAG1
MAG2
Resolution (Lp/mm)
Brivo OEC 715/785
≥
2.0
≥
2.6
≥
3.1
Brivo OEC 865
≥ 2.2
≥
3.0
≥
3.5
If the resolution check fails, refer to section 4.6.8 for Camera Focus calibration.
3.6.5.
Dose Limitation Check
Normal Fluoro Dose rate check
1.
Operate C-Arm to make the generator being top and the image intensifier being bottom.
2.
Position the detector of dosimeter at the center of the X-ray beam 30cm above the grid of image
intensifier. Cover the image intensifier with lead plate thicker than 3mm.
3.
Set KVp as 110kVp, increase mA to maximum in manual mode. Measure the dose rate while
making exposure. Limits vary by the dose group, which was configured in software configuration
section by country.
4.
Refer to below table verify the corresponding Dose group according to country’s Dose group
requirement (refer to 5.2.4).
Continuous Fluoro:
≤
47.5 mGy/min (Dose group: Normal 50 and HLF 50)
≤
47.5 mGy/min (Dose group: Normal 50 and HLF 100)
≤
83.6 mGy/min (Dose group: Normal 88 and HLF 100)
HLF Dose rate check:
1.
Position the dosimeter in the center of the X-ray beam 30cm above the grid. Cover the image
intensifier with lead thicker than 3mm for protection of the image intensifier.
2.
Set kVp as 110kVp, increase mA to maximum in manual mode. Measure the dose rate while
making exposure. Limits vary by the dose group, which was selected in software configuration
section by country.
3.
Refer to below table verify the corresponding Dose group according to country’s Dose group
requirement (refer to 5.2.4).
Continuous Fluoro:
≤
47.5 mGy/min (Dose group: Normal 50 and HLF 50)
≤
95 mGy/min (Dose group: Normal 50 and HLF 100)
≤
95 mGy/min (Dose group: Normal 88 and HLF 100)
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...