Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-14
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
please call for service.
0115004
Warning
Flash write fault
during beam
alignment
Calibration data save
exception. The mainframe
stopped. Please restart the
system. If this issue persists,
please call for service.
1.MCB may be abnormal
0115005
Warning
Flash erase
fault during
Collimator
calibration
Calibration data save
exception. The mainframe
stopped. Please restart the
system. If this issue persists,
please call for service.
1.MCB may be abnormal
0115006
Warning
Flash write fault
during DAP
calibration
Calibration data save
exception. The mainframe
stopped. Please restart the
system. If this issue persists,
please call for service.
Flash write fault during DAP
calibration
1.MCB may be abnormal
0117001
Warning
KV is not
accurate
System execution exception.
Please try again. If this issue
persists, please call for service.
1. Redo the kV calibration.
2. Check the bundle cable
3. Check emergency switch
4. Check the MCB.
5. Check the kV control
board.
6. Replace the monoblock.
0117002
Warning
MA is not
accurate
System execution exception.
Please try again. If this issue
persists, please call for service.
1. Redo the mA calibration.
2. Check the bundle cable
3. Check Hand/ foot switch
signal and emergency
switch.
4. Check the MCB.
5. Check the FDB.
7. Replace the monoblock.
0117003
Warning
MA is very low
(<0.05ma)
System execution exception.
Please try again. If this issue
persists, please call for service.
1. Redo the mA calibration.
2. Check the bundle cable
3. Check emergency switch.
4. Check the MCB.
5. Check the FDB.
6. Replace the monoblock.
0117004
Warning
MA is very
high(>23.5ma)
System execution exception.
Please try again. If this issue
persists, please call for service.
Redo mA calibration
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...