Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
4-22
4.6.
Beam Alignment and Image Calibration
4.6.1.
Camera Centering
Camera Centering Check:
1.
Let C-arm in Image Intensifier Up/Monoblock Down position, Log in software service mode, in
calibration Tab, click “Beam alignment…” button.
2.
Rough (recommend to 1.8LP/mm) adjust camera focus to see the light image margin clearly, refer to
4.6.8.
3.
Perform Camera Centering Check as below:
Fail image:
ABS and ABC mode there is black
margin in circular mask ring
Fail image:
None_ABS and None_ABC,
manual exposure in 60KV, 1.1mA with
contrast in 65 and rightness in 40, there is
black margin in circular mask ring
Fail
Fail
Circular Mask
Circular Mask
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...