Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
9-14
9.5.
Radiation Safety Specification
Operators are strongly urged to be familiar with the current recommendations from the International
Commission on Radiological Protection. (ICPR)
Take radiation protection of the patient and operator into consideration seriously.
Scatter data in this manual is only for reference.
9.5.1. Scatter Data
Patients must be kept as far as possible from the focus, with minimum distance of 20cm.
Make sure not to operate when a plastic cover in x-ray assembly is disassembled, which
properly make the distance meet the above requirements.
The operating table matched this equipment should be made of small attenuation
equivalent material, otherwise, X-ray dose may be increased.
Filter in the X-Ray tube assembly
3.35mm Al
Additional filter
0mmAl
Total filter (without removable filter)
3.35mmAl
Leakage radiation
Dose area product rate is less than
1.0mGy/h, averaged over any area of
100cm
P
2
P
of which no principal linear
dimension exceeds 20cm at 1m from the
focal spot under conditions of loading.
(110kVp, 2mA)
SID (focal spot to image receptor distance)
100cm
WARNING
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...