Replacement
7-57
Do not touch the capacitors unless power has been completely removed and they have
been discharged.
2.
Disconnect the cables. Release the secured screw and remove the capacitor.
3.
Replace a new one. Reassemble all the parts.
4.
Power on the system and take exposure. Verify that the system can work normally.
Resonant Capacitor (P/N: 5329281) and DC bus fuse kit (P/N: 5075999)
1.
Remove the rear cover of the C-Arm. Release the 3 screws and open the package box.
2.
Disconnect the cables connected to the resonant capacitor or DC bus fuse. Loose the screws and
replace the new one, remove the DC bus fuse label and paste the new one.
3. Power on the system and take exposure. Verify that the system can work normally.
DC bus fuse
Resonant Capacitor