Replacement
7-63
Connectors List:
Connector
To
To 5085781
To 5085708 (connector on 5085781)
PE
GND
PJ3
Power Distribution Unit Board
PJ6
Power Distribution Unit Board
Ethernet
Cable
Computer
RS-485
Computer
Replacement procedure:
1.
Remove Workstation back, side, front covers refer to section 7.3.4.
2.
Remove all cable pressures.
3.
Cut all able ties, and unplug the all connecters.
4.
Assemble the new cable and route it according to original.
Keep new cable routed as original to ensure the package of workstation can be open and
closed smoothly.
5.
Upgrade the Interconnect cable socket (if Rev.6 or lower version) with the hexagon socket screws
.
6.
Install all the cable pressures.
7.
Recheck all connectors are correctly connected and cables are routed well.
8.
Power on the system, and take an exposure.
9.
If system works well, turn off system and assemble all covers.
NOTE
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...