Troubleshooting
6-23
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
restart the system. If this issue
persists, please call for service.
0204505
Warning
Load font
resource failed
Font not installed correctly.
Text on left monitor may be
abnormal. Please restart the
system. If this issue persists,
please call for service.
make sure Torch font
installed correctly
0204506
Warning
Load icon file
failed
System Icon files loading
failure. Some icons may be
missed. Please restart the
system. If this issue persists,
please call for service.
check if icon file crashed
0204507
Error
Allocate
memory failed
Run out of memory. Please
restart the system. If this issue
persists, please call for service.
check memory using state
Application
0306501
Warning
Load
connectivity
XML failed
Connectivity configuration
loading failure. Some
configuration may be missed.
Please restart the system. If
this issue persists, please call
for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0306502
Warning
Get text
resource failed
Get text resource failure.
Application display may be
abnormal. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0306503
Warning
Enconter
undefined
exception
Undefined exception. Please
restart the system. If this issue
persists, please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0306504
Warning
Init
AbandonReaso
n.xml file failed
Abandon Reason loading
failure. Abandon Exam
function will be impacted.
Please restart the system. If
this issue persists, please call
for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...