Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
3-20
3.
Lay UPS on side as below picture and remove the battery cover (battery cover is under UPS).
4.
Pull out the battery of UPS and plug the power supply cable into the power pin of UPS box.
5.
Push back battery and install the UPS cover.
6.
Reassemble UPS.
The UPS receives AC power whenever the system power switch is on. The battery
recharges if necessary. If the AC main power fails, the UPS supplies power to computer
using its internal battery. The PDU will sounds “bi-bi-bi-bi”.
Do not connect any other devices to the UPS receptacles. UPS battery should be
disconnected before transportation. Battery can be left in the UPS.
WARNING