Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-20
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
install software.
0201504
Warning
Open camera
failed
Image acquisition exception.
System can’t take exposure.
The mainframe stopped.
Please restart the system. If
this issue persists, please call
for service.
1. Check connections
between camera
and workstation.
2. Check ETH1 port on
workstation mainboard.
3. Replace the camera.
0201505
Warning
Load driver
configuration
file failed
Camera driver exception.
System can’t take exposure.
The mainframe stopped.
Please restart the system. If
this issue persists, please call
for service.
1. Check connections
between camera
and workstation.
2. Check ETH1 port on
workstation mainboard
0201506
Warning
Get driver
thread failed
Camera driver exception.
System can’t take exposure.
The mainframe stopped.
Please restart the system. If
this issue persists, please call
for service.
End capture failed.
1. If Gige camera is not
working correctly, please
change camera.
2. If LED on ethnet interface
is not twinkle, please insert
the network cable into
ethnet interface again or
replace another network
cable.
3. If all above methods can’t
fix the exception, please re-
install software.
0201507
Warning
Set capture
thread priority
failed
Image acquisition exception.
System can’t take exposure.
The mainframe stopped.
Please restart the system. If
this issue persists, please call
for service.
End capture failed.
1. If Gige camera is not
working correctly, please
change camera.
2. If LED on ethnet interface
is not twinkle, please insert
the network cable into
ethnet interface again or
replace another network
cable.
3. If all above methods can’t
fix the exception, please re-
install software.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...