Brivo OEC 715/785/865/865 Mobile C-Arm X-Ray Product Service Manual
2-60
Signals from sealing board to MCB:
Name
Description
Attribution Scale Comment
OIL_TEMP
Output voltage of
temperature sensor
Analog
0-5V -
OIL_TEMP_GND
Reference of OIL_TEMP
Ground
0V
-
THERMAL_SWITCH_A
One termination of
temperature switch
Passive
-
THERMAL_SWITCH_A &
THERMAL_SWITCH_B
connected means that
temperature is lower than
threshold.
THERMAL_SWITCH_A &
THERMAL_SWITCH_B
disconnected means that
temperature is higher than
threshold.
THERMAL_SWITCH_B
The other termination of
temperature switch
Passive
-
MA_CAT
The cathode of mA sample
Analog
-
-
KV_GND
The anode of mA sample
Analog
-
-
Technical Specifications:
Nominal high voltage
110 kV
Nominal input power
2.2 kW
Maximum anode heat content
53 kJ
Target material
Tungsten
Target angle
10
°
Small focal spot (IEC336) (Normal Fluoro, High
Level Fluoro, Digital Spot)
0.6
×
1.4 (IEC336)
Large focal spot (IEC336) (Radiography)
1.4 (IEC336)
Filtration equiv.
0.55 mm Al
Anode
Stationary
Typical filament specification
4.2 V, 3.5A
Maximum filament current
4.6A
Weight
0.65 kg
Nominal heat capacity
76kHU
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...