Calibration
4-5
4.2.2.
kV Adjustment
Provide appropriate X-ray protection on surface of Image Intensifier in calibration and
verification to protect Image Intensifier.
1.
Remove the C-arm console cover. Check the connection of IGBT board cables.
2.
Remove the cover of E-Box. Install MCB board with extend board. Remove the J46 jumper on
extend board. Connect the probes of ampere meter to TP222 (+) and TP210 (-) on extend board.
3.
Power on the system. Check that DS9, DS10 and DS3 indicators on kV control board light.
4.
Hold down keys “Ctrl +Shift + Alt +
→
” to invoke the service tool. Login service tool with password.
Click Calibration Tab, then click “mA Calibration” button, “CA11” will display on control panel.
5.
Rotate R40 and R27 on kV control board clockwise to the end. (If applicable)
6.
Test the frequency of TP4 on kV Control Board with Oscilloscope. Adjust R13 on kV Control Board
to make the frequency of TP4 is 80kHz
±
1kHz (TP6 and TP9 are GND.)
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...