Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-46
7.3.12.
Monitors
To prevent monitor power lost, a binding is used to fix the monitor power plug (see the
picture in step6). The binding is provided with the monitor FRU. Make sure the binding is still
used after the monitor replacement.
Perform the procedures below to replace the monitors.
1.
For Brivo OEC865 need remove monitor back
covers.
2.
Loosen the screw and remove the monitor
receptacle access cover. Cut the tying strip on
power cable and pull out the two cables. Make
mark on cables to identify which monitor they
should connect to.
3.
Remove the sheet metal support of the exposure indicator.
For Brivo OEC 715/785 need remove the strip cover of the exposure indicator wiring firstly.
Brivo OEC 715/785
Brivo OEC 865
4.
Loosen the screws of the block that secure the
cables. And then loosen the screws on the
sheet metal support and remove it.
5.
Release the four screws on the monitor and remove the monitor.
CAUTION
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...