Interface:
AC power input
The AC power input is listed as follows.
Name
Description
Pin
Attribution
Scale
Comment
MF_AC28V_L
AC power input for motor
J1-1
POWER
28V AC
----
MF_AC28V_N
AC power input for motor
J1-3
POWER
28V AC
----
Up/down orders interface
Signals used to control the motion direction of the lift column motor are listed as follows.
Name
Description
Pin
Attribution
Scale
Comment
UP_N
Up order from the up key
J2-1
Digital
0V/24V
LOW =the up key is pressed
GND_PS1
----
J2-3
Ground
0V
----
DOWN_N
Down order from the
down key
J2-5
Digital
0V/24V
LOW =the down key is pressed
DOWN_SECURITY
_N
Down security order from
the down security key
J2-6
Digital
0V/24V
LOW=the down security key is
pressed
Key switch interface
The key switch signals to enable or disable the motion of the lift column motor are listed as follows.
Name
Description
Pin
Attribution
Scale
Comment
KEY_N
Disable motion signal from
the key switch.
J3-1
Passive
Short/open
short=Enable motion
GND_PS1
0V.
J3-2
Ground
0V
----
Lift column motor interface
The DC power supplies for the motion of the lift column motor are listed as follows.
Name
Description
Pin
Attribution
Scale
Comment
MOTO1
DC power to the lift column motor.
Up motion when the voltage is
higher than MOTO2’ s.
J4-1
Power
0--+40VDC
The DC voltage between
MOTO1 and MOTO2 shall be
within the range: -40VDC--
+40VDC
MOTO2
DC power to the lift column motor.
Down motion when the voltage is
higher than MOTO1’ s.
J4-4
Power
0--+40VDC
Mainframe Control Board interface
The signals to MCB are listed as follows.
Name
Description
Pin
Attribution
Scale
Comment
DOWN_PROTECT_T
O_MCB_N
Proximity protecting signal
to MCB.
J5-1
Digital
0V/+15V
LOW=Proximity protect
UDB_OVERCUR_N Overcurrent signal to MCB.
J5-2
Digital
0V/+15V
LOW =Error, overcurrent
UP_TO_MCB_N
Up signal to MCB.
J5-3
Digital
0V/+15V
LOW =The motor is at
up motion
DOWN_TO_MCB_N Down signal to MCB.
J5-5
Digital
0V/+15V
LOW =The motor is at
down motion
DOWN_SECURITY_
TO_MCB_N
Down security key signal to
MCB.
J5-6
Digital
0V/+15V
LOW =Down security
key is pressed
UDB_ENABLE
Preserved signal from MCB
to disable the motor motion.
J5-7
Digital
0V/+15V
LOW =Disable motion
+24V_UDB
DC 24V power input to UDB.
J5-8
Power
+24V
----
GND_PS1
----
J5-9
Ground
0V
----
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...