Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-56
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0407508
Warning
Enumerate
devices failed
Enumerate devices failure.
Please restart the system. If
this issue persists, please call
for service.
System error: can’t
enumerate devices.
0407512
Warning
Run makeisofs
script failed
Create CD/DVD file exception.
Please retry or insert another
disk. If this issue persists,
please call for service.
Make sure the makeisofs
script is the exact version
that is shipped with
product.
0407514
Warning
Run synccopy
script failed
Copy to USB disk exception.
Please check the USB disk or
replace with another disk. If
this issue persists, please call
for service.
Make sure the synccopy
script is the exact version
that is shipped with
product.
0407516
Error
Start
application
failed
System module init failure.
Please restart the system. If
this issue persists, please call
for service.
Make sure there is enough
memory and the udev
library is correctly installed.
0407517
Error
Report ready
failed
System start failed,
MediaManagement report
ready failed. Please restart the
system. If this issue persists,
please call for service.
Application can’t be booted,
contact service.
0407519
Warning
Run command
failed
System command execution
failed. Please try again. If this
issue persists, please call for
service.
System error. Call to system
function failed.
0407521
Warning
Print head
temperature
exception
Print head temperature
exception. Please restart the
local printer and retry. If this
issue persists, please call for
service.
Check the temperature for
the printer.
0407522
Warning
Printer paper
feed error
Paper feed error. Please take
the paper out of printer and
put it in again. If this issue
persists, please call for service.
Get the paper out of printer
and put it in.
0407523
Warning
Printer is not
initialized
Printer is not ready. Please
make sure the printer is well-
connected and already turned
on. And then try again. If this
issue persists, please call for
service.
Make sure the printer has
been initialized.
Summary of Contents for Brivo OEC 715
Page 2: ......
Page 19: ...Chapter1 Introduction and Safety...
Page 41: ...Introduction and Safety 23 46 54 20 18 26...
Page 55: ...Chapter2 System Overview...
Page 137: ...Chapter3 Installation...
Page 212: ...Chapter4 Calibration...
Page 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Page 284: ...Chapter5 Software...
Page 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Page 335: ...Chapter6 Troubleshooting...
Page 408: ...Chapter7 Replacement...
Page 418: ...Replacement 7 11 166...
Page 488: ...Chapter8 Periodic Maintenance...
Page 502: ...Periodic Maintenance 8 15...
Page 505: ...Chapter9 Technical Reference...
Page 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Page 526: ...11 Appendix System Schematics...